electrograms (Abbott Laboratories)
Structured Review

Electrograms, supplied by Abbott Laboratories, used in various techniques. Bioz Stars score: 90/100, based on 1 PubMed citations. ZERO BIAS - scores, article reviews, protocol conditions and more
https://www.bioz.com/product/electrogram/electrograms/pmc07749220-42-0-1
Average 90 stars, based on 1 article reviews
Images
1) Product Images from "Treatment of symptomatic bradycardia due to T-wave oversensing with implantation of a new generator incorporating delayed decay and threshold start sensitization algorithms"
Article Title: Treatment of symptomatic bradycardia due to T-wave oversensing with implantation of a new generator incorporating delayed decay and threshold start sensitization algorithms
Journal: HeartRhythm Case Reports
doi: 10.1016/j.hrcr.2020.06.024
Figure Legend Snippet: Postimplant Abbott (Abbott Park, IL) electrograms (auto-gained right ventricular signal). Prior to programming, there is intermittent overcounting of the sensed T wave (top pane) at VVI 30 beats per minute (bpm). Following programming of decay delay at 220 ms and threshold start at 1.0 mV of the paced beat there is elimination of T-wave oversensing at both 40 bpm and 130 bpm.
Techniques Used:
Related Articles